THE HT9 Modular Patient Monitor of Hwatime is an advanced modular monitor designed to provide comprehensive and customizable monitoring according to clinical needs. Patient compatible adults, pediatrics and neonates, It combines precision, speed and flexibility thanks to its oscillometric technology and multiple optional modules.
🔹 Non-invasive blood pressure measurement (NIBP)
-
Technology : oscillometric during inflation
-
Measurement range:
-
Adult/Pediatric: 40–240 mmHg
-
Newborn: 40–150 mmHg
-
-
Typical measurement cycle: < 40 s
-
Selectable cycle modes: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 min
-
STAT mode: continuous playback for 5 minutes
-
Specific beaches:
-
Adult: SYS 40–270 / DIA 10–215 / Mean 20–165 mmHg
-
Pediatric: SYS 40–135 / DIA 10–100 / Mean 20–110 mmHg
-
Newborn: SYS 40–200 / DIA 10–150 / Mean 20–165 mmHg
-
-
Maximum permissible pressure:
-
Adult: 300 mmHg
-
Child: 240 mmHg
-
Newborn: 150 mmHg
-
-
Resolution : 1 mmHg
-
Precision : ±5 mmHg
🔹 Temperature
-
Channels: 2
-
Measurement range: 0–50 °C
-
Precision : ±0.1 °C
-
Display resolution: 0.1 °C
-
Compatible probes: YSI 400 and YSI 700 Series
🔹 Calculations and data storage
-
Calculations available: dose, ventilation, renal function, hemodynamics, oxygenation
-
Storage capacity:
-
Data: 200 hours
-
Alarm groups: 1000
-
NIBP History: 720 hours
-
Curves: 48 hours
-
🔹 Optional modules
-
CO₂ (main / lateral)
-
Range: 0–15 %
-
Resolution: 1 mmHg
-
Accuracy: < 50 ml/min
-
Respiratory rate: 2–120 bpm
-
-
IBP (invasive blood pressure)
-
Channels: 2
-
Precision : ±2 % or ±1 mmHg
-
Measurable pressures: ART, PA, CVP, RAP, LAP, ICP, P1, P2
-
Beaches:
-
PA: -6 to 120 mmHg
-
CVP/RAP/LAP/ICP: -10 to 40 mmHg
-
P1/P2: -10 to 300 mmHg
-
-
🔹 Recorder (optional)
-
Kind : 50mm thermal printer
-
Resolution : high definition
-
Speed : 25 mm/s or 50 mm/s
-
Channels: 3







Reviews